- SKU: incubadora-infantil-servo-digital-modelo-pilli-1000-sd-tactil-7
- BARCODE: incubadora-infantil-servo-digital-modelo-pilli-1000-sd-tactil-7
- VENDOR: Prosalud.me
TEHSA PILLI 1000 SD Digital Servo Infant Incubator, 7" Touchscreen, Cat. TEH-PSDT
Taxes included.
Shipping calculated at checkout.
Couldn't load pickup availability
Infant servo digital incubator, PILLI 1000 SD 7" touch screen model, Cat. TEH-PSDT Brand TEHSA
Description:
Medical device with a canopy or closed cover to help control the air and skin temperature of neonate or newborn patients.
Standard accessories:
- Inlet port for oxygen supply of 15 Kg/cm2; Air outlet of 15 Kg/cm2.
- Air filter that offers resistance to dust. Not reusable. Maximum life of 3 months. High efficiency (0.8 µm) of the filter in stopping contaminants in the air intake.
- Filling pipe for water supply to generate a humid environment inside the incubator.
- IV hook
Optional accessories:
- Electric height adjustment (option for pedal and memory control)
- Integrated phototherapy with timer and hour counter. Integrated scale.
- Power backup to keep the control module on.
- Oxygen servo control.
- Oximetry (SpO2 and Bpm).
Control system:
- Professional control system based on a latest-generation microprocessor that allows high-precision and safe operation for the care of critically ill patients.
- Control and programming system for operating parameters via a user-friendly 7” color touch screen.
- Digital display of all thermal control parameters, monitored parameters, and heater power.
- Digital control for setting the control temperature and selecting the operating mode.
- Self-test system when the equipment is turned on and during operation.
- Air temperature servo control.
- Skin temperature servo control.
- Relative humidity servo control.
- Heating power modulation in 3 stages.
Alarms (audible and visual):
- High temperature.
- System / sensor failure.
- Ventilation failure.
- Power failure.
- Includes alarm silence control.
Operating modes:
- AIR: Control of the air temperature circulating inside the canopy.
- SKIN OR SERVOCONTROLLED: The skin temperature of the neonate in the hepatic area (of vital organs) is controlled by means of the skin sensor and the equipment control.
Mattress:
- Hypoallergenic, antistatic, and resistant to temperatures above 40°C.
- System with the possibility of three positions:
- Horizontal.
- Trendelenburg.
- Reverse Trendelenburg.
- Radiolucent mattress base to facilitate X-ray taking.
Base:
- Metal frame cabinet.
- Mobile with doors and shelves to allow the placement of equipment, clothing, and medications.
- Braking system on 2 of its 4 wheels.
Dimensions:
- Height: 134 cm + - 1 cm.
- Width: 56 cm + - 1 cm.
- Length: 94 cm + - 1 cm.
- Weight: 72 Kg. + - 1 Kg.
Operating data:
- Selection range in air mode: 28.0°C to 38.0°C
- Selection range in skin mode: 36.0°C to 38.0°C
- Thermometer resolution: 0.1°C
- High temperature alarm: 39.0°C + - 0.1°C.
- Accuracy of the indicated temperature: + - 0.2°C