Introduction
ECG300G is an electrocardiograph that can collect 12-lead ECG signals and print the waveform using a thermal printing system. It features recording and displaying ECG waveform in AUTO/Manual mode, automatically measuring and diagnosing ECG waveform parameters, indicating "Lead off" and "Paper lack", multilingual interface, AC/DC, selection of any rhythm lead, case database management.
Function
1) 12-lead synchronization for ECG collection, adopts digital signal processing technology and obtains high-quality ECG waveform through power frequency filter, baseline filter and EMG filter of the ECG signal.
2) Display of 3/6/12 leads ECG, printing mode, sensitivity, paper speed and filter status, on one screen etc., convenient for contrastive diagnosis.
3) Multiply print modes and formats, including manual, auto 4 × 3, auto 3 × 4 + 1, auto 3 × 4, auto 2 × 6 + 1, auto 2 × 6, auto 3-2 + 1, auto 3-2, Auto 1 × 12 + 1, Auto 1 × 12, rhythm 4, rhythm 3 and rhythm 2, etc. The length of printed waveform can be adjusted and with the time print function, which meets different requirements.
4) With auto-analysis and auto-diagnosis functions for routine ECG parameters, it provides measurement results and auto-diagnosis conclusion for HR, P-R interval, P Duration, QRS Duration, T Duration, Q-T interval, Q-Tc, P Axis, QRS Axis, T Axis, R(V5), S(V1), R(V5)+ S(V1), etc. which reduces the burden of the physician.
5) Built-in memory stores up to 1000 cases, convenient for case review and statistics.
6) Multi-language interface and report (Chinese, English, Spanish, Turkish, Polish, Italian, French, German, Portuguese, Kazakh, Russian, Ukrainian, Serbian and Slovenian).
Performance
-
Input mode: floating and defibrillation protection
-
Frequency response: 0.05Hz ~ 150Hz(-3dB ~ + 0.4dB)
-
CMRR: >60dB, >100dB (Add filter)
-
Time constant: ≥ 3.2S
-
Patient leakage current: <10 & micro;A
-
Calibration voltage: 1mV
-
Sensitivity: 2.5, 5, 10, 20, 40mm/mV ± 5%, standard sensitivity: 10mm/mV ± 2%
-
Noise level: <15 & micro;Vp-p
-
Input circuit current: ≤ 50nA
-
Input impedance: ≥ 50MΩ
-
Sampling accuracy: 12 bits
-
Safety classification: Class I, Type CF and defibrillation-proof applied part
-
Recording mode: thermal printing system
-
Auto record: set the record according to auto record format and mode, automatically switching leads, measuring and analyzing.
-
Rhythm record: set the record according to rhythm record format and mode, automatically measuring and analyzing.
-
Manual record: set the record according to record format, manually switching leads.
-
Measurement parameters: HR, P-R interval, P Duration, QRS Duration, T Duration, Q-T interval, Q-Tc, P Axis, QRS Axis, T Axis, R(V5), S(V1), R(V5)+ S(V1)
-
Paper size: 80mm(W)× 20m(L)
-
Sampling frequency: 1000Hz
-
EMG interference filter: 25Hz/35Hz (-3dB)
-
Power frequency filter: AC 50Hz/60Hz (-20dB)
-
Paper speed:
-
Auto: 25, 50 mm/s, ± 5%
-
Rhythm: 25, 50 mm/s, ± 5%
-
Manual: 5, 6.25, 10, 12.5, 25, 50 mm/s, ± 5%
Accessories
Standard:
One Lead cable
One limb electrode
One chest electrode
One thermal recording paper
One power cord
One ground wire
One user manual
ECG Sync software + USB cable
Physical characteristics
-
Power supply
-
AC: 100V ~ 240V(50/60Hz)
-
DC: 7.4V/3500mAh rechargeable lithium battery
-
Fuse specification: two AC time-lag fuses (φ 5 × 20mm), T1.6AL250V
-
Waterproof degree: IPX0
-
Working mode: continuous work